The StatStrip Lactate Hospital Meter System is a handheld point-of-care testing system that brings lactate testing directly to the patient’s bedside.
Lactate is currently measured on blood gas analyzers, which creates numerous problems for bedside testing. Blood gas analyzers require anaerobic arterial samples, large sample volumes (100-200 microliters), and long analysis times (up to 2.5 minutes).
In addition, blood gas analyzers are complex to operate, fixed in location, and expensive to purchase and use for serial lactate testing.
StatStrip Lactate provides the fastest turnaround time possible (13 seconds), on the smallest whole blood sample (0.6 microliters), and can be easily operated by medical and nursing staff. StatStrip Lactate is a very low-cost device that makes lactate testing practical and affordable in any size ED, ICU, or medical unit. It is fully connectable to a hospital's IT system through NovaNet.
Weight: 0.49 lb (220 g)
StatStrip Xpress Lactate not yet available in the U.S. and Canada
Designed for applications that do not need data interfacing and connectivity, Xpress Lactate retains all of the analytical performance features of StatStrip Lactate.
Weight: 0.2 lb (75 g)
Size: 3.6 in x 2.3 in x 0.9 in (91 mm x 58 mm x 23 mm)
Data Storage:
Patient & QC Tests: 400 tests total (FIFO)
Connectivity:
Data Transfer: Strip Port Connection to USB
Data Program: Nova Microsoft-Excel based data transfer software
Battery Information:
Type: 3V Li Button Battery
Features: Replaceable
Life: Minimum 600 tests
Additional Features:
Tests:
Tests Measured: Blood Lactate, Hematocrit Corrected
Test Reported: Lactate
Test Time: 13 Seconds
Test Strip Volume: 0.6 µL
Test Methodology: Electrochemistry
Sample Types & Operating Modes:
Whole Blood: Arterial, Venous
Lactate Measurement Range:
0.3-20.0 mmol/L (3-180 mg/dL)
No Interferences From:
Hematocrit, Ascorbic Acid, Uric Acid, Acetaminophen (Paracetamol), Bilirubin
Operating Ranges:
Temperature: 59°F-104°F (15°C- 40°C)
Altitude: Up to 15,000 feet
Humidity: 10% to 90% relative humidity
Hematocrit: 20% – 65%
Reagents and Strips:
Strips: 25 strips per vial; 2 vials per package
QC: Two levels, sold separately
Linearity: Four levels available
Test Strip & QC Stability:
24 months from date of manufacturing
3 months open-vial stability
Certifications and Compliance: Nova Biomedical is certified to FDA Quality System Regulations and EN ISO 13485:2016 Complies to IVDD Tested according to: EN 61010-1:2010, EN 61010-2-101:2015, EN 60825-1/A1:2014.
CSA Standards. Patented Technology. Patents 6,287,451 / 6,837,976 / EP 1 212 609 / CA 2,375,092 / EP 1497 449 / JP 405 0078
Specifications current as of revision date.